Welcome. Your work on Trifluoroacetate counterion was a useful addition — it was one of the thinner pages in its topic area.
One convention to note for next time: numeric claims need an inline citation each, not one citation at the end of the paragraph, because the paragraph usually ends up split later and the citation then attaches to the wrong sentence. See Project:Sourcing guidelines. — ToleranceTove (talk) 20:12, 15 December 2025 (UTC)
Gradient conditions in Reference standard
The purity figure quoted in Reference standard is given without the gradient or the column chemistry it was obtained on. Two laboratories running different gradients can resolve a close-eluting impurity differently and both be right, so the conditions are part of the result rather than a footnote to it. Do you have them from the source? — ColdChainCleo (talk) 18:35, 19 July 2026 (UTC)
Good point, and I do not have them — the figure came from a certificate that does not state the method. I have said so rather than implying it was measured under known conditions. — TechnicianTeal (talk) 04:40, 20 July 2026 (UTC)