94The vial-yield arithmetic follows from the wastage figure. Each dose consumes 0.20 mL delivered plus 0.070 mL retained, or 0.270 mL. From 2.00 mL, seven full doses can be drawn (7 × 0.270 = 1.89 mL), with 0.11 mL remaining — insufficient for an eighth. The nominal ten doses have become seven, a 30% reduction. With a fixed-needle insulin syringe at 0.005 mL, each dose consumes 0.205 mL and nine full doses can be drawn, with 0.155 mL remaining.{{r|ppdeadspace}}94The vial-yield arithmetic follows from the wastage figure. Each dose consumes 0.20 mL delivered plus 0.070 mL retained, or 0.270 mL. From 2.00 mL, seven full doses can be drawn (7 × 0.270 = 1.89 mL), with 0.11 mL remaining — insufficient for an eighth. The nominal ten doses have become seven, a 30% reduction. With a fixed-needle insulin syringe at 0.005 mL, each dose consumes 0.205 mL and nine full doses can be drawn, with 0.155 mL remaining.{{r|ppdeadspace}}
9595
+96| Device dead space | Volume consumed per 0.20 mL dose | Full doses from 2.00 mL | Loss against nominal |
+97|---|---|---|---|
+98| 0.005 mL | 0.205 mL | 9 | 10% |
+99| 0.020 mL | 0.220 mL | 9 | 10% |
+100| 0.070 mL | 0.270 mL | 7 | 30% |
+101| 0.100 mL | 0.300 mL | 6 | 40% |
+102
96=== Priming and air management ===103=== Priming and air management ===
97The mechanism by which the dose error arises is the presence of gas where liquid is assumed. Descriptions of injection technique in manufacturers' [[Prescribing information|prescribing information]] and in nursing references accordingly treat expulsion of air before the dose is measured as a defined step, and treat the graduation as valid only for a gas-free barrel and hub.{{r|who2015}}104The mechanism by which the dose error arises is the presence of gas where liquid is assumed. Descriptions of injection technique in manufacturers' [[Prescribing information|prescribing information]] and in nursing references accordingly treat expulsion of air before the dose is measured as a defined step, and treat the graduation as valid only for a gas-free barrel and hub.{{r|who2015}}