1{{Infobox concept1{{Infobox concept
2| name = Syringe dead space2| name = Syringe dead space
+3| image = syringe.svg
+4| caption = Dead space lies distal to the plunger tip at full depression: the hub cavity, the needle lumen and any recess in the plunger face.
3| Also known as = Residual volume; hold-up volume; wastage volume5| Also known as = Residual volume; hold-up volume; wastage volume
4| Location = Needle hub, needle lumen, plunger-tip recess6| Location = Needle hub, needle lumen, plunger-tip recess
61| 31G | 0.261 mm | 0.133 mm | 0.18 µL |63| 31G | 0.261 mm | 0.133 mm | 0.18 µL |
6264
+65Volumes are computed as the product of cross-sectional area and length; outside diameters follow the metric equivalents of the Birmingham gauge series used in ISO 9626, and inside diameters are typical regular-wall values.{{r|iso9626}} A 27G half-inch needle therefore holds 0.44 µL, roughly one part in 160 of a 70 µL hub. Changing needle gauge or length has almost no effect on dead space; changing the hub design changes it by an order of magnitude.
+66
63=== Effect of needle length ===67=== Effect of needle length ===
64Doubling cannula length from 12.7 mm to 25.4 mm doubles the lumen volume, taking a 25G needle from 0.67 to 1.34 µL. Against a hub of 60–70 µL this is immaterial. Needle length is selected on anatomical grounds — 4, 6 and 8 mm pen needles and 12.7 mm syringe needles are the common lengths for [[Subcutaneous injection|subcutaneous administration]] — and not on dead-space grounds.{{r|iso7864}}68Doubling cannula length from 12.7 mm to 25.4 mm doubles the lumen volume, taking a 25G needle from 0.67 to 1.34 µL. Against a hub of 60–70 µL this is immaterial. Needle length is selected on anatomical grounds — 4, 6 and 8 mm pen needles and 12.7 mm syringe needles are the common lengths for [[Subcutaneous injection|subcutaneous administration]] — and not on dead-space grounds.{{r|iso7864}}
78[[Category:Dosing and titration]]82[[Category:Dosing and titration]]
79[[Category:Preparation and handling]]83[[Category:Preparation and handling]]
+84[[Category:Clinical practice]]